Applications
Volcanic Gases Analysis
Analyzer Description
- Configuration:
- 5-valves, 2-packed columns, 2-micropacked columns, 2-capillary columns, methanizer, TCD, TCD, FID, injector oven, nitrogen purge, custom low pressure sampling system
- Sample type:
- Volcanic gas, prepared as per journal article: “The Magmatic and Hydrothermal Dominated Fumarolic System at the Active Crator of Lascar Volcano, Northern Chile”
- Compounds Separated:
- Helium, hydrogen, carbon dioxide, argon, oxygen, nitrogen, methane, carbon monoxide
- Concentration Range:
- 50 ppm – 20% for helium and hydrogen
- 200 ppm – 100% for carbon dioxide and nitrogen
- 200 ppm – 0.2% for argon and oxygen
- 200 ppm – 50% for methane
- 0.5 ppm – 5% for carbon monoxide
- Methods met:
- F. Tassi, F. Aguilera, O. Vaselli, E. Medina, D. Tedesco, A. Delgado Huertas, ..S. Kojima (2009). The magmatic- and hydrothermal-dominated fumarolic system at the Active Crater of Lascar Volcano, Northern Chile. Bull Volcanol 71:171-183
- Wasson-ECE W4376, W4365
Typical Chromatogram
Key Features and Benefits
- Sample loops are purged with nitrogen between analyses.
- Argon and oxygen separation is achieved without cryogenics.
- A vacuum gauge allows for direct measurement of sample loop pressure for a more accurate calculation of moles injected.