Vinylidene Fluoride Impurities

Analyzer Description

Configuration:
3-valves, 2-capillary columns, methanizer, FID, MSD
Sample type:
Vinylidene fluoride
Compounds Separated:
Tetrafluoroethylene, ethylene, 1-chloro-1,1-difluoroethane, carbon monoxide, carbon dioxide
Lower detection limits:
1 ppm
Methods met:
Wasson-ECE W0785

Key Features and Benefits

  • Approximately a 20 minute analysis time.
  • The mass spectrometer quantifies the analytes based on the integrated signal for only the target ions listed allowing identification and quantitation of components by ion extraction even if retention times overlap.
  • Specially deactivated tubing and valves ensure repeatability and the lowest detection limits.
  • Instrument can be configured with additional analysis for impurities in tetrafluoroethylene and impurities in hexafluoropropylene.