ASTM D2163 – Analysis of Liquefied Petroleum (LP) Gases and Propene Concentrates

Analyzer Description

Configuration:
FID, 3-valves, 2-capillary columns
Sample type:
Liquefied petroleum (LP) gases and mixtures of propane and propene, excluding high-purity
propene
Compounds Separated:
  • methane
  • ethane
  • ethylene
  • propane
  • cyclopropane
  • propylene
  • acetylene
  • isobutene
  • propadiene
  • n-butane
  • trans-2-butene
  • 1-butene
  • isobutylene
  • cis-2-butene
  • neopentane
  • isopentane
  • methyl acetylene
  • n-pentane
  • 1,3-butadiene
  • 3-methyl-1-butene
  • trans-2-pentene
  • 2-methyl-2-butene
  • 1-pentene
  • 2-methyl-1-butene
  • cis-2-pentene
  • neohexane
  • n-hexane
  • n-heptane
  • benzene
  • C8+ backflush
Lower detection limits:
10 ppm for individual analytes
80 ppm for composite backflush
Methods met:
ASTM D2163

Typical Chromatogram

FID Chromatogram

Key Features and Benefits

  • Approximately a 15 minute analysis time.
  • Analyzer can be configured dual train to double the sample throughput.
  • Analyzer can be configured for both liquid and gas sample analysis.
  • The backflush peak can be tailored to specific samples (e.g. C6+ or C8+ backflush composite peaks).
  • Gas chromatograph can be configured to contain additional applications.