Applications

Coal and Biomass Fuel Gas Impurities

Analyzer Description

Configuration:
FID, MSD, SCD, 3-valves, 4-capillary columns, deans switch
Sample type:
Coal and biomass fuel gas
Compounds Separated:
Hydrogen sulfide, carbonyl sulfide, methyl mercaptan, ethyl mercaptan, carbon disulfide, benzene, toluene, ethyl benzene, xylenes, phenol, cresol, polycyclic aromatics, acetylene, ethane, ethylene, propane, propylene, iso-butane, n-butane, 1-butene, iso-butylene, dimethyl sulfide, thiophene, benzothiophene
Lower detection limits:
10 ppb for sulfur compounds
1 ppb for remaining compounds
Methods met:
Wasson-ECE W1420

Key Features and Benefits

  • Fused silica lined tubing ensures repeatability and the lowest detection limits.
  • Sulfur treated inlet and sample lines ensure no sulfur lost during sample introduction.
  • A 3-way decoking valve alleviates any coking problems that might occur in the SCD burner during the oxidation process and ensures maximum response and consistency.
  • The backflush peak can be tailored to specific samples (e.g. C6+ or C8+ backflush composite peaks).